Evaluation of the needle positioning accuracy of a light puncture robot under MRI guidance: results of a clinical trial on healthy volunteers

sept. 1, 2018·
Ghelfi Julien
,
Moreau-Gaudry Alexandre
,
Hungr Nikolaï
,
Fourd Céline
,
Veron Baptiste
,
Medici Maud
,
Chipon Émilie
,
Cinquin Philippe
,
Bricault Ivan
· 0 min de lecture
Architecture diagram
Résumé
Purpose To assess the accuracy of Light Puncture Robot (LPR) as a patient-mounted robot, in positioning a sham needle under MRI guidance for abdominal percutaneous interventions. Materials and Methods This monocentric, prospective and non-controlled study was approved by the ethics review board. The study evaluated the accuracy of LPR V3 to achieve a virtual puncture in 20 healthy volunteers. Three trajectories were tried on each volunteer, under 3-T MRI guidance. Results Accuracy under 5 mm in attaining a 10 cm-deep target was reached in 72% of attempts after 2 robot motions with a median error of 4.1 mm [2.1; 5.1]. Median procedure time for one trajectory was 12.9 min [10.2; 18.0] and median installation time was 9.0 min [6.0; 13.0]. Conclusion LPR accuracy in the deployment of a sham needle inside the MRI tunnel and its setup time are promising. Further studies need to be conducted to confirm …
Type
Publication
Cardiovascular and interventional radiology, vol 41 no 9